Skip to main content
. 2013 Jun 3;12:181. doi: 10.1186/1475-2875-12-181

Table 2.

Limit of detection for pLDH in three rapid diagnostic tests by using artificially diluted samples of Plasmodium vivax

pLDH (ng/mL) Parasitaemia(/μL) SD Bioline (n = 5) OptiMAL (n = 5) Humasis(n = 5)
5,000
8,700.0
100%
100%
100%
500
870.0
100%
100%
100%
250
435.0
100%
100%
100%
50
87.0
100%
100%
100%
25
44.0
100%
0%
100%
5
8.7
0%
0%
0%
0.5
0.9
0%
0%
0%
0 0 0% 0% 0%