Published in final edited form as: Phys Med Biol. 2009 Sep 24;54(20):6065–6078. doi: 10.1088/0031-9155/54/20/003

Table 2.

Peak neutron doses for hybrid brass/plastic collimators with overall thickness sufficient to stop 235 MeV protons. The geometry used here (figure 1, no pre-collimator) is identical to those for the calculations shown in figure 3. Typical statistical uncertainties (1σ) in the results, related to the Monte-Carlo simulation, are ±0.3%.

Brass thickness (mm) Plastic thickness (mm) Total collimator thickness (mm) Collimator-related peak neutron dose (pGy/incident proton)
65 0 65 0.36
55 25 80 0.29
50 41 91 0.25
45 59 104 0.22
40 75 115 0.20
35 94 129 0.18
25 128 153 0.17
0 195 195 0.16