Abdominal MDCT Standard-Dose Protocols Used in the Prospective Trial
Protocol |
|||
---|---|---|---|
Parameter | IV Contrast Enhanced | Urolithiasis | Supine CT Colonography |
Scanner | Discovery CT750 HDa | Discovery CT750 HDa | Discovery CT750 HDa |
Scan type | Helical | Helical | Helical |
Rotation time (s) | 0.5 | 0.8 | 0.5 |
Beam collimation (mm) | 40 | 40 | 40 |
No. of detector rows | 64 | 64 | 64 |
Pitch | 0.516 | 0.516 | 0.984 |
Speed (mm/rotation) | 20.64 | 20.64 | 39.36 |
Detector configuration | 64 × 0.625 | 64 × 0.625 | 64 × 0.625 |
Slice thickness for noise index (mm) | 1.25 | 1.25 | 1.25 |
Scan FOV | Large body | Large body | Large body |
Peak kilovoltage | 120 | 120 | 120 |
Smart mA (GE Healthcare) range (mA) | 60–660 | 40–660 | 30–300 |
Noise index | 24 | 28 | 50 |
Reconstruction (filtered back projection) | |||
Displayed FOV (cm) | 36–50 | 36–50 | 36–50 |
Reconstruction type | Standard | Standard | Standard |
Window width/level (HU) | 400/50 | 400/50 | 400/50 |
Reconstruction option | Plus | Plus | Plus |
Slice thickness (mm) | 2.5 | 2.5 | 2.5 |
Interval (mm) | 1.5 | 1.5 | 1.5 |
Note—The protocols considered standard dose in this study are based on use of 40% adaptive statistical iterative reconstruction, which affects the relative dose reduction seen at low dose. The specific protocol for the accompanying low-dose series was derived by adjusting the noise index–slice thickness pairing (and tube current range) to allow targeted 70–90% dose reduction (by projected dose-length product) relative to the standard-dose series.
GE Healthcare.