Table 4.

Implementation of FMEA in analytical phase

Failure mode Effect SI Cause PI Control measure DI RPI Proposed action
Malfunction of reagent Useless result 9 Expired 3 Check expiration date 1 27 None
Malfunction of reagent Useless result 9 NC storage temperature 6 Visual check of reagent 10 540 Add temperature monitoring system
Malfunction of reagent Useless result 9 Contaminated 8 QC before run 10 720 Add QC after run
Malfunction of calibrator Calibration failure 8 Expired 3 Check expiration date 2 48 None
Malfunction of calibrator Calibration failure 8 NC storage temperature 6 Visual check of calibrator 2 96 None
Malfunction of calibrator Calibration failure 8 Contaminated 8 Visual check of calibrator 2 128 Freeze single doses and use once

SI severity index, PI probability index, DI detectability index, RPI risk probability index