Table 1.
Consensus Error Grid Analysis and Statistical Parameters of the Pooled Data Pairs of Six Patients with Abdominal and Upper Arm Placement of the FiberSensea
FiberSense upper arm | FiberSense abdomen | |
n | 1144 | 1142 |
Zone A | 94.6% | 83.8% |
Zone B | 5.4% | 16.2% |
Zones C–E | 0.0% | 0.0% |
MARD | 8.3% | 11.4% |
Median ARD | 6.5% | 8.4% |
Data are pooled over 14 days of wearing time.