Table 3.
Consensus Error Grid Analysis and Statistical Parameters of the Pooled Data Pairs of Six Patients with Abdominal and Upper Arm Placement of the FiberSense Continuous Glucose Monitoring at 2, 7, and 14 days of Wearing Time
| n | MARD | Median ARD | Zone A | Zone B | |
| FiberSense upper arm | |||||
| 2 days | 206 | 8.9% | 7.2% | 92.7% | 7.3% |
| 7 days | 173 | 6.4% | 5.0% | 98.8% | 1.2% |
| 14 days | 178 | 8.9% | 6.8% | 92.7% | 7.3% |
| FiberSense abdomen | |||||
| 2 days | 206 | 9.6% | 7.1% | 86.9% | 13.1% |
| 7 days | 205 | 13.0% | 10.8% | 83.9% | 16.1% |
| 14 days | 161 | 10.6% | 5.6% | 81.4% | 18.6% |