Table 2.

Hypoglycemia Detection Performance Measures of Electrocardiogram and Electroencephalogram Featuresa

Feature name True positive rate False positive rate Detection time median Detection time range
ECG features
QTc 6/9 1/9 8 min 0–26 min
TpTec 5/9 2/9 8 min 0–23 min
HR 5/9 3/9 7 min 2–16 min
EEG features
EEG delta power 3/9 0/9 8 min 8–13 min
EEG theta power 8/9 0/9 13 min 4–24 min
EEG alpha power 4/9 0/9 14 min 7–20 min
a

The detection time is the time from when significant changes of the features are detected to the termination of experiments based on the clinical status of the patient (see text).