Skip to main content
. 2013 Jul;20(7):1084–1088. doi: 10.1128/CVI.00316-13

Table 1.

Comparison of BioPlex 2200 Syphilis IgG, ARCHITECT Syphilis TP, and TPHA to IgG WB assay

Assay and result IgG WB result (no. of samples)
% sensitivity (95% CI) % specificity (95% CI) % agreement (95% CI) κ value
Positive Negative
BioPlex 2200 Syphilis IgG
    Positive 352 48 100 (99–100) 89.7 (86.6–92.3) 94.2 (92.5–95.8) 0.88
    Negative 0 420
ARCHITECT Syphilis TP
    Positive 352 101 100 (99–100) 78.4 (74.4–82.1) 87.7 (85.4–89.9) 0.76
    Negative 0 367
TPHA
    Positive 351 1 99.7 (98.5–99.9) 99.8 (98.8–100) 99.8 (99.4–99.9) 0.98
    Negative 1 467