Skip to main content
. 2013 Jul;51(7):2396–2399. doi: 10.1128/JCM.00151-13

Table 1.

Xpert MTB/RIF versus AFB smear microscopy and culture results for 79 clinical specimensa

Xpert MTB/RIF result Microscopy and culture results
Smear-positive specimens (n = 19)
Smear-negative specimens (n = 60)
All specimens (n = 79)
No. of cultures positive for MTC No. of cultures positive for NTM No. of negative cultures No. of cultures positive for MTC No. of cultures positive for NTM No. of negative cultures No. of cultures positive for MTC No. of cultures positive for NTM No. of negative cultures
M. tuberculosis detected 14 0 0 1 0 2c 15 0 2
M. tuberculosis not detected 0 5 0 3b 2 50 3 7 50
Not interpretable 0 0 0 0 0 2 0 0 2
Total 14 5 0 4 2 54 18 7 54
a

Among the 79 specimens were 71 respiratory specimens (57 sputa, 10 bronchial aspirates, and 4 bronchoalveolar lavage fluids) and 8 nonrespiratory specimens (3 biopsies, 1 CSF, 1 ascitic fluid, 1 pleural fluid, 1 pus, and 1 pharynx aspirate).

b

All nonrespiratory specimens.

c

Both were respiratory specimens.