. 2013 Jul;51(7):2195–2203. doi: 10.1128/JCM.00622-13

Fig 4.

Fig 4

(A) Distribution of Amplicor CT/NG test OD450 values, showing (i) the most-recent Amplicor CT equivocal zone, (ii) the earlier Amplicor equivocal zone, and (iii) the broad equivocal zone identified using the observations of this study. (B) Amplicor CT test OD450 (x axis) against C. trachomatis plasmid quantity (y axis), as determined by ddPCR. OD450 values of 0.1, 0.8, and 3.1 are indicated. ddPCR values of zero are assigned an arbitrarily low (0.001 plasmids/μl) value. Positive ddPCR with high estimates of the infectious load are infrequent when the Amplicor CT/NG OD450 is <3.1.