Table 1.

Percent CT number accuracy and percent noise reduction observed in five clinical cases. Both CT number and standard deviation are measured in a ROI (20 × 20 pixels) in the relative smooth liver region, such as those shown in Figures 4 and 5. ACNR: anticorrelated noise reduction [12].

Dataset Basis material Algorithm Mean density (mg/cc) Standard deviation Absolute density error (mg/cc) % Density error % Noise reduction
Case 1 
(Figure 4)
Water Original 29.16 72.22
ACNR 26.00 24.26 3.16 10.8% 66.4%
Proposed 27.85 30.29 1.31 4.5% 58.1%
Iodine Original −961.41 29.71
ACNR −961.22 9.45 0.19 0.002% 68.2%
Proposed −961.33 10.32 0.08 0.008% 65.3%

Case 2 
(Figure 5)
Water Original 15.28 88.27
ACNR 24.25 27.99 8.97 58.7% 68.3%
Proposed 15.42 28.17 0.14 0.9% 68.0%
Iodine Original −967.92 31.42
ACNR −967.41 9.83 0.51 0.05% 68.7%
Proposed −966.31 10.98 0.61 0.2% 65.1%

Case 3 Water Original 5.74 85.39
ACNR 6.81 25.99 1.07 18.6% 69.6%
Proposed 5.93 27.07 0.19 3.7% 68.3%
Iodine Original −934.50 34.77
ACNR −935.87 9.26 1.37 0.2% 73.2%
Proposed −935.61 9.91 1.11 0.2% 71.5%

Case 4 Water Original 9.42 45.94
ACNR 7.48 15.89 1.94 20.6% 65.4%
Proposed 8.70 16.57 0.72 7.6% 63.9%
Iodine Original −993.72 12.84
ACNR −994.56 3.41 0.84 0.08% 73.4%
Proposed −994.25 3.94 0.53 0.05% 69.3%

Case 5 Water Original 12.98 161.68
ACNR 15.81 48.47 2.83 21.8% 70.0%
Proposed 13.85 51.20 0.87 6.7% 68.3%
Iodine Original −990.19 20.36
ACNR −989.32 6.25 0.87 0.09% 69.3%
Proposed −988.89 6.76 1.3 0.1% 66.8%

Original
Average ACNR 2.18 13.1% 69.3%
Proposed 0.69 2.40% 66.5%

Original
Maximum ACNR 8.97 58.7% 73.4%
Proposed 1.31 7.6% 71.5%