Overview of the limitations of the temperature measurement techniques in gas turbines. Taken from Feist et al. [22].

Method Disadvantage/Limitation
Thermocouples Intrusive
Limited number
Costly installation
Bonding to ceramic surfaces
Electromagnetic Interference
Difficult to use on rotating components
Thermal Paints Intrusive
Time consuming (calibration and experiment)
Irreversible
Discrete values and poor resolutionCostly
Pyrometry Sensitive to stray light
Changes in emissivity
Translucency of ceramic coatings
Cleanness of optics
Thermographic Phosphor Decreasing signals with increasing temperatures
Bonding of the phosphor
Phosphor coating can be intrusive