Skip to main content
. 2008 Sep 25;8(9):5942–5960. doi: 10.3390/s8095942

Table 1.

Characteristics of ECL sensors based on the [Ru(bpy)3]2+/TPA system.

Sensor Platform LOD, M Linear Range, M RSD, % Ref.
Silica/Eastman-AQ55D/ [Ru(bpy)3]2+ 1 × 10-7 2 × 10-5 – 1 × 10-3 1.9 5
TiO2/Nafion/[Ru(bpy)3]2+ 1 × 10-7 1 × 10-7 – 1 × 10-3 3.9 4
V2O5/Nafion/[Ru(bpy)3]2+ 1 × 10-8 5 × 10-8 – 1 × 10-3 2.5 9
ZrO2/Nafion (on 3D Au structure)/ [Ru(bpy)3]2+ 5 × 10-10 1 × 10-9 – 1 × 10-5 0.74 8
Pt NPs/Eastman-AQ55D / [Ru(bpy)3]2+ 1 × 10-15 0.6 21
Fe3O4 NPs/Nafion/[Ru(bpy)3]2+ 5 × 10-8 1 × 10-7 – 1 × 10-3 3.9 27
Fe3O4 NPs/Silica/[Ru(bpy)3]2+ 6.5 × 10-9 6.9 × 10-8 – 7.3 × 10-4 0.5 26
SNPs/[Ru(bpy)3]2+ 1 × 10-8 2.6 × 10-8 – 1.3 × 10-3 5.2 20
[Ru(bpy)3]2+ SNPs/chitosan 2.8 × 10-9 8.5 × 10-9 – 8.1 × 10-5 28
[Ru(bpy)3]2+ SNPs/CNTs 2.8 × 10-9 8.5 × 10-9 – 7.9 × 10-4 29
Nafion/CNTs/[Ru(bpy)3]2+ 1 × 10-9 3 × 10-9 – 1 × 10-4 <10 6
PSP/CNTs/[Ru(bpy)3]2+ 6 × 10-9 38
Nafion/CNTs/TiO2/[Ru(bpy)3]2+ 1 × 10-8 5 × 10-8 – 1 × 10-3 <4 39