Skip to main content
. Author manuscript; available in PMC: 2013 Jul 9.
Published in final edited form as: Pediatr Crit Care Med. 2011 Nov;12(6):643–648. doi: 10.1097/PCC.0b013e31821926a5

Table 1.

True positive (TP) and false positive (FP) alarms with alarm thresholds at 60 mg/dL and no forward prediction, and at 70 mg/dL with 15 minute forward prediction. At both alarm settings a TP was defined as an alarm that occurred with a confirmed blood glucose values less than or equal 60 mg/dL and a FP was defined as an alarm with a confirmed blood glucose value greater than or equal to 60 mg/dL. Sensor Enabled Therapy Intervention (SETI) was defined as a sensor alarm that resulted in an exogenous glucose bolus being given in advance of BG falling below 60 mg/dL. Sensor Enabled Therapy Alert (SETA) was defined as a sensor alarm resulting in an exogenous glucose bolus not given in time to prevent BG from falling below 60 mg/dL.

Subjects (N) BG<40
mg/dL
BG 40–60
mg/dL
TP FP %
detected
SETI SETA
Alarm set at 60 mg/dL
All (64) 3 19 0 9 0 0 0
TGC (30) 3 17 0 9 0 0 0
STD (34) 0 2 0 3 0 0 0
Alarm set at 70 mg/dL
All (247) 3 46 20 334 41% 8 11
TGC (125) 2 38 18 262 45% 8 8
STD (122) 1 8 3 74 33% 0 3