Skip to main content
. 2013 Jun 25;8:153. doi: 10.1186/1748-717X-8-153

Table 3.

Impact on setup technique on D95, D05 and Dav of the PTV as compared to the original treatment plan

PTV Laser setup
CBCT setup
D95 D05 Dav D95 D05 Dav
mean
−12.3%
−3.9%
−5.9%
−0.4%
−0.5%
−0.1%
max
−31.8%
−17.7%
−20.4%
−0.8%
−3.9%
−0.4%
min −1.7% −0.2% −0.5% 0.2% 0.2% 0.4%

Changes are given in % of the respective dose value of the treatment plan, negative values represent a dose reduction and positive values an increase.