Table 1.

Accuracies of the system used to determine setup margins for perpendicular-to-beam (a) and beam-axis (b) directions

(a) Perpendicular to beam direction (lateral)
Displacement of beam axis ≦±0.5 mm
Position of RC ≦±0.5 mm
Position of patient collimator ≦±0.5 mm
Position of MLC ≦±0.5 mm
Displacement of ICs for horizontal and vertical beam courses ≦±0.5 mm
X-ray beam axis from IC ≦±0.5 mm
FPD axis from IC ≦±0.5 mm
Positioning system (image registration) (ΔP) ≦±2 mm
Machining accuracy of RC ≦±0.3 mm
Machining accuracy of patient collimator ≦±0.3 mm
(b) Beam axis direction
CT conversion (2 %) for 15 cm (if 10 cm, ± 2 mm) ≦±3 mm
Range ≦±1 mm
Machining accuracy of RC ≦±0.3 mm
Uniformity of RC material (1 %) for 5 cm ≦±0.5 mm