Skip to main content
. Author manuscript; available in PMC: 2014 Jun 21.
Published in final edited form as: Phys Med Biol. 2013 May 17;58(12):4099–4118. doi: 10.1088/0031-9155/58/12/4099

Figure 10.

Figure 10

The optimal swale-depth sd* can be reduced by decreasing magnification and by choosing x-ray tube and detector that minimize xt and xd, respectively. Assuming a 5 mm mechanical clearance for the x-ray tube, xt = 52.5 mm and 37 mm, were obtained from Boone (Boone, 2004) and from specifications of M-1500 x-ray tube (Varian Medical Systems), respectively. Assuming a similar clearance for the detector, xd = 39.2 mm and 23.6 mm, were obtained from specifications of the PaxScan® 4030CB and PaxScan® 4030MCT, respectively (Varian Medical Systems). Computations were performed assuming LM =5 mm. The arrow labeled A indicates sd* for the system geometry in (Boone, 2004) and the arrow labeled B indicates sd* for a recently installed system at our institution.