Published in final edited form as: J Magn Reson Imaging. 2010 Jan;31(1):255–261. doi: 10.1002/jmri.22019

Table 1. Comparisons of contrast to noise ratio (CNR) for CODEA MRA/MRV acquired at 7T vs 3T.

The blood vessels are indicated by arrows in Fig. 4, with the sizes of blood vessels corresponding to the thickness of the arrows. The CNR values are given as mean ± standard deviation. The values within the parentheses in the “CNR at 7T” column indicate those from CODEA MRV without phase-mask filtering.

Blood vessels CNR at 7T CNR at 3T Ratio (7T / 3T)
Artery 1 (“large”) 218±55 108±14 2.0
Artery 2 (“medium”) 116±10 72±12 1.6
Artery 3 (“small”) 159±44 51±8 3.1
Vein 1 (“large”) 73±4 (66±5) 42±6 1.7
Vein 2 (“medium”) 77±10 (69±8) 36±3 2.1
Vein 3 (“small”) 67±5 (61±7) 31±2 2.2