Determined values for f and KSV of branched fiber model and of evanescence wave application Limit of Detection (LOD) were determined by linear fit of 0 mbar-100 mbar, calculating with Henry constant (0.0013 mol/kg·bar [27]) resulting LOD of oxygen in water in mol/L.

Branched Fiber Arrangement Evanescent Wave Application


30 m 90 m 20 m 80 m
f 0.8 ± 0.1 0.8 ± 0.1 0.8 ± 0.1 0.8 ± 0.1
KSV/mbar1 0.014 ± 0.003 0.013 ± 0.002 0.012 ± 0.002 0.014 ± 0.002
LOD/*105M (H2O) 0.3 ± 0.2 1.5 ± 0.2 2 ± 0.5 2.3 ± 0.2