Comparison of the bioMérieux EasyQ KPC with a routine, phenotypic method of KPC detection

EasyQ KPCa No. of specimens with routine method result
% (95 CI)b
Positive Negative Total Sensitivity Specificity PPV NPV
All specimens 93.3 (77.9–99.2) 99.0 (98.0–99.6) 77.8 (60.9–89.9) 99.7 (99.1–100)
    Positive 28 8 36
    Negative 2 768 770
    Total 30 776 806
Stools 90.0 (55.5–99.8) 98.3 (95.1–99.7) 75.0 (42.8–94.5) 99.4 (96.9–100.0)
    Positive 9 3 12
    Negative 1 174 175
    Total 10 177 187
Swabs 95.0 (75.1–99.9) 99.2 (98.1–99.7) 79.2 (57.9–92.9) 99.8 (99.1–100.0)
    Positive 19 5 24
    Negative 1 594 595
    Total 20 599 619
a

Final EasyQ result, after retests to resolve invalid results.

b