1
|
INITIALIZE
|
Initialize hardware
|
2
|
TRAPF18
|
Trap [18F]fluoride for 120 s using 3 psi inert gas
|
3
|
ELUTEF18
|
Flow eluent through elute path for 120 s using 3 psi inert gas
|
4
|
EVAPORATE
|
Evaporate reactor 1 at 110°C for 300 s using 15 psi inert gas
|
5
|
ELUTEF18
|
Flow MeCN-1 through elute path for 90 s using 3 psi inert gas
|
6
|
EVAPORATE
|
Evaporate reactor 1 at 110°C for 120 s using 10 psi inert gas
|
7
|
ADD
|
Add MeCN-2 to reactor 1
|
8
|
EVAPORATE
|
Evaporate reactor 1 at 110°C for 120 s using 10 psi inert gas and cool to 30°C
|
9
|
ADD
|
Add mannose triflate to reactor 1
|
10
|
REACT
|
React reactor 1 at 130°C for 300 s and cool to 35°C with an additional 120s cooling delay
|
11
|
EVAPORATE
|
Evaporate reactor 1 at 110°C for 120 s using 10 psi inert gas
|
12
|
ADD
|
Add HCl to reactor 1
|
13
|
REACT
|
React reactor 1 at 130°C for 300 s and cool to 35°C with an additional 120-s cooling delay
|
14
|
TRANSFER
|
Transfer the contents of reactor 1 through the purification cartridge to the collection vial. Push with 10 psi inert gas for 30 s
|
15
|
ADD
|
Add Water-1 to reactor 1
|
16
|
TRANSFER
|
Transfer the contents of reactor 1 through the purification cartridge to the collection vial. Push with 10 psi inert gas for 30 s
|
17
|
ADD
|
Add Water-2 to reactor 1
|
18 |
TRANSFER |
Transfer the contents of reactor 1 through the purification cartridge to the collection vial. Push with 10 psi inert gas for 45 s |