Table 1.

Simulated values of back pressure in cm H2O generated in the desiccant mouthpiece for varying porosities and particle sizes. Simulation is for 5 grams of calcium chloride.

Particle size Porosity
0.25 0.35 0.45 0.55 0.65
0.35 mm 243.50 66.68 22.46 8.23 3.01
0.70 mm 60.91 16.67 5.61 2.05 0.75
1.05 mm 27.07 7.41 2.49 0.91 0.33
1.40 mm 15.22 4.17 1.40 0.52 0.18