Table I.

Measured distance of brain to CFOV in an actual experiment versus design intent in IndyPET-II (2nd and 3rd columns) and microPET P4 (4th and 5th columns).

IndyPET-II microPET P4
Design Intent (mm) Actual measurement (mm) Design Intent (mm) Actual measurement (mm)
Radial distance from each brain to CFOV 30 34 35 37
Axial distance from each brain to CFOV 4 5 10 15
Total Distance 30.3 34.4 36.4 40