Table 1.

Performance characteristics of HIV RDTs

RDT Sensitivity% (n = 150)
Specificity% (n = 150)
Manufacturer’s SEN* SEN Manufacturer’s SPE* SPE (95% CI) NPV% (95% CI) PPV% (95% CI)
Determine 100 100 99.9 85.2 (77.4, 91.1) 100 (96.3, 100) 67.3 (52.9, 79.7)
STAT-PAK 100 100 100 99.1 (95.3, 99.9) 100 (96.8, 100) 97.2 (85.5, 99.9)
Uni-Gold 100 100 99.7 97.4 (92.6, 99.5) 100 (96.8, 100) 92.1 (78.6, 98.3)
First Response 100 100 99.2 97.4 (92.6, 99.5) 100 (96.8, 100) 92.1 (78.6, 98.3)
Advanced Quality 100 100 100 100 (96.8, 100) 100 (96.8, 100) 100 (90.0, 100)

PPV = positive predictive value; NPV = negative predictive value; CI = confidence interval; RDT = rapid diagnostic test; SEN = sensitivity; SPE = specificity

*

Manufacturers’ test characteristic performance (kit inserts) on serum; PPV and NPV dependent on the prevalence of population tested

Evaluation findings