Table 3. Factors influencing variability in vascular SO2 by retinal oximetry.

Co-variate SD of SO2 Measurements Relative Effect Size* p-value
Between-subject variability Un-IR group vs. IR group 2.23% vs. 3.63% 1.63 0.14
Between- vessel variability (within the same patient) Un-IR vs. IR group 5.06% vs. 8.19% 1.62 <0.001
Arterioles vs. venules 2.78% vs. 5.06% 0.55 <0.001
Vessel width 0.993 0.71
Between- vessel variability (within the same patient) Un-IR vs. IR group
Arterioles vs. venules
Vessel width
Same vessel variability (repeatability of measurement) Un-IR vs. IR group 2.39% vs. 2.48% 1.04 0.41
Arterioles vs. venules 1.70% vs. 2.39% 0.71§ <0.001
Vessel width 1.01 0.27

SD: Standard deviation; Un-IR: unirradiated group; IR: irradiated group.

*

Standard deviation of SO2 measurements in one patient co-variate category divided by another co-variate category.

Between-vessel SD in the irradiated group is 1.62 times the unirradiated group for vessels of the same type and width.

Between-vessel SD for arterioles is 0.55 that of venules for vessels of same width and in the same group of patients.

§

The SD of repeated measurements of the same vessel is 0.71 fold smaller for arterioles than for venules in vessels of the same width and in the same group of patients.