Table 2.

Calibration factors for converting OSL readings to thimble chamber readings, together with the measuring conditions. For each exposure, the body bowtie filter was used, and the beam collimation was 20 mm at the iso-center.

kVp Bowtie PMMA (cm) HVL (mm Al) Calibration factor Std. error R2
80 Body 4 6.0 1.14 0.0063 0.9998
100 Body 0 6.5 1.21 0.0098 0.9997
100 Body 2 6.9 1.24 0.0037 1.0000
100 Body 4 7.3 1.26 0.0090 0.9997
120 Body 0 7.5 1.34 0.0102 0.9997
120 Body 2 7.9 1.36 0.0086 0.9993