Published in final edited form as: IEEE Trans Ultrason Ferroelectr Freq Control. 2012 Dec;59(12):2777–2784. doi: 10.1109/TUFFC.2012.2519

TABLE III.

Temperature Dependence of the Performance of the Pb(Mg1/3Nb2/3)–PbTiO3 (PMN-PT) Linear-Array Ultrasonic Transducer.

Temperature (°C) Echo Vpp (mV) fc (MHz) −6-dB BW (%) f1 (MHz) f2 (MHz) IL (dB)
21.5 69.83 3.36 57.14 2.10 4.32 27.65
40 69.28 3.40 64.71 2.30 4.50 28.62
60 55.78 3.29 61.80 2.27 4.30 29.51
80 54.74 3.30 60.61 2.30 4.30 30.20
100 46.75 3.53 71.95 2.26 4.80 31.04
120 51.36 3.21 61.68 2.22 4.20 30.27
110 30.88 3.24 53.09 2.38 4.10 33.40
160 10.41 3.05 47.87 2.32 3.78 42.90

BW = bandwidth; IL = insertion loss.