TABLE III.
conventional cosine TMS | cTMS (C = 716 μF) | ||||||||
---|---|---|---|---|---|---|---|---|---|
| |||||||||
VCo (V) | C‡(μF) | tp (μs) | ΔWC (J) |
|
tp (μs) | ΔWC (J) |
|
||
| |||||||||
1,327 | 63 | 49 | 56 | 165 | 28 | 43 | 54 | ||
1,028 | 121 | 66 | 64 | 244 | 38 | 48 | 80 | ||
882 | 183* | 81 | 71 | 329 | 45 | 49 | 96 | ||
810 | 240 | 92 | 79 | 408 | 50 | 52 | 112 |
All configurations are matched to produce same amount of neuronal membrane depolarization for membrane time constant of τm = 150 μs.
total energy storage capacitance C = C1 + C2.
approximate Magstim 200 configuration.