Table 2.

Concordance, sensitivity, specificity, and accuracy of Sanger sequencing, real-time polymerase chain reaction with TheraScreen® kit, pyrosequencing, and chip array infiniti® assay measured using 454 next-generation sequencing as the reference method

Method Kappa of concordance Sensitivity Specificity Accuracy
Sanger sequencing 0.791 34/40 (85%) 100% 54/60 (90%)
TheraScreen (ΔCt) 0.857 36/40 (90%) 100% 56/60 (93%)
Pyrosequencing (percent mutated) 0.892 37/40 (93%) 100% 57/60 (95%)
Chip array assay (ratio)* 0.891 36/39 (92%) 100% 56/59 (95%)

Note:

*

One sample not assessed, values calculated on 59/60 cases.