Skip to main content
. Author manuscript; available in PMC: 2013 Dec 18.
Published in final edited form as: Anal Chem. 2012 Dec 5;84(24):10549–10553. doi: 10.1021/ac302997n

Table 1.

Slope of each trendline showing FIQ Correction Factor/μM SNAPtide calculated for each temperature. These slopes can be used to calculate FIQ Correction Factors at SNAPtide concentrations between 0.00–10.0 μM and at reaction temperatures between 25–50°C if the 0.0 μM SNAPtide sample is assumed to have an FIQ Correction Factor of 1.00(Calculation 1). Standard deviations of the mean for the triplicate samples are shown.

Temperature 25°C 30°C 37°C 40°C 45°C 50°C
FIQ Correction Factor/μM SNAPtide Slope −0.0264 ± 0.0013 −0.0255 ± 0.0017 −0.0247 ± 0.0007 −0.0232 ± 0.0022 −0.0255 ± 0.0015 −0.0258 ± 0.0036