TABLE 1.

Protocol Specifications and Typical Technical Chest Radiographic Parameters Used in the National Lung Screening Trial

Parameter Film-Screen Computed Radiography Digital Radiography

Tube voltage (kV) 120–150 100–140 110–150
Maximum skin entrance exposure (mGy)a 0.3 0.4 0.3
Maximum exposure time (ms) 40 40 40
Source-image distance (in)a ≥ 72 ≥ 72 ≥ 72
Antiscatter device (grid) ≥ 10:1 at 103 lines/in (stationary) or 80 lines/in (reciprocating) Optimal for system Optimal for system
Minimum collimation To image receptor To image receptor To image receptor

Note—72 in = 182.88 cm, 1 in = 2.54 cm.

a

Skin entrance exposure may exceed these guidelines in large individuals.