TABLE 1.
Parameter | Film-Screen | Computed Radiography | Digital Radiography |
---|---|---|---|
| |||
Tube voltage (kV) | 120–150 | 100–140 | 110–150 |
Maximum skin entrance exposure (mGy)a | 0.3 | 0.4 | 0.3 |
Maximum exposure time (ms) | 40 | 40 | 40 |
Source-image distance (in)a | ≥ 72 | ≥ 72 | ≥ 72 |
Antiscatter device (grid) | ≥ 10:1 at 103 lines/in (stationary) or 80 lines/in (reciprocating) | Optimal for system | Optimal for system |
Minimum collimation | To image receptor | To image receptor | To image receptor |
Note—72 in = 182.88 cm, 1 in = 2.54 cm.
Skin entrance exposure may exceed these guidelines in large individuals.