TABLE II.

Contrast-To-Noise Ratio (CNR) of the Lesions on Shear Velocity Images and B-Mode Images.

Treatment parameters (pulses/location) CNR in agar phantoms
CNR in ex vivo kidneys
Shear wave velocity image B-mode image Shear wave velocity image B-mode image
0 (control) 0.1 ± 0.1 0.1 ± 0.1 0.1 ± 0.1 0.1 ± 0.1
100 6.0 ± 1.1 0.5 ± 0.1 1.2 ± 0.6 0.2 ± 0.2
200 6.8 ± 1.2 0.7 ± 0.1 1.8 ± 0.8 0.3 ± 0.2
300 7.3 ± 1.4 0.9 ± 0.1 2.4 ± 0.8 0.4 ± 0.2
500 8.1 ± 1.0 1.1 ± 0.1 2.6 ± 0.9 0.6 ± 0.2
1000 8.9 ± 2.8 1.3 ± 0.1 3.2 ± 0.9 0.8 ± 0.2
1500 8.2 ± 3.3 1.3 ± 0.1 3.1 ± 1.2 0.9 ± 0.2
2000 5.3 ± 2.8 1.4 ± 0.1 3.0 ± 1.2 1.0 ± 0.3

The mean ± standard deviation is listed (N = 9 for phantom studies; N = 8 for ex vivo tissue studies).