TABLE II.
Treatment parameters (pulses/location) | CNR in agar phantoms
|
CNR in ex vivo kidneys
|
||
---|---|---|---|---|
Shear wave velocity image | B-mode image | Shear wave velocity image | B-mode image | |
0 (control) | 0.1 ± 0.1 | 0.1 ± 0.1 | 0.1 ± 0.1 | 0.1 ± 0.1 |
100 | 6.0 ± 1.1 | 0.5 ± 0.1 | 1.2 ± 0.6 | 0.2 ± 0.2 |
200 | 6.8 ± 1.2 | 0.7 ± 0.1 | 1.8 ± 0.8 | 0.3 ± 0.2 |
300 | 7.3 ± 1.4 | 0.9 ± 0.1 | 2.4 ± 0.8 | 0.4 ± 0.2 |
500 | 8.1 ± 1.0 | 1.1 ± 0.1 | 2.6 ± 0.9 | 0.6 ± 0.2 |
1000 | 8.9 ± 2.8 | 1.3 ± 0.1 | 3.2 ± 0.9 | 0.8 ± 0.2 |
1500 | 8.2 ± 3.3 | 1.3 ± 0.1 | 3.1 ± 1.2 | 0.9 ± 0.2 |
2000 | 5.3 ± 2.8 | 1.4 ± 0.1 | 3.0 ± 1.2 | 1.0 ± 0.3 |
The mean ± standard deviation is listed (N = 9 for phantom studies; N = 8 for ex vivo tissue studies).