Table 2.
Analyzer | Instruments | Type | Resolution | Mass accuracy | Dynamic Range |
---|---|---|---|---|---|
Quadrupole | QQQ QToF |
Beam | 1 – 2 K | ~ 1 ‰a | 5 – 6 |
Ion trap | LIT | Trapping (Electric field) | 1 – 2 K | ~ 1 ‰a | 3 – 4 |
ToF | QToF | Beam | 10 – 50 K | 5 – 10 ppm | 4 |
Orbitrap | FT | Trapping (Electric field) | 7.5 – 240 K | 500 ppb – 10 ppm | 4 |
ICR | FT | Trapping (Magnetic and axial DC fields) | 100 – 500 K | ~100 ppb | 3 |
parts-per-thousand.