Skip to main content
. 2012 Nov 8;1:45. [Version 1] doi: 10.12688/f1000research.1-45.v1

Table 1. Clinical alarm scenarios that were programmed into the bedside monitors.

Detected Scenarios Parameters Limits/Trigger Time
(scenario name) (detect what?) (maximum of four) (lower & upper violation
for x seconds or relative triggers
in % over a defined time
in sec/min)
SVT + BP onset of paroxysmal atrial fibrillation HR (Pulse)
ART sys
Pulse (HR)
+40% within 59 sec
-15% within 59 sec
>110 bpm for 20 sec
Vtach + BP Vtach with low blood pressure HR (Pulse)
PVC
ARTsys
Pulse (HR)
+30 bpm within 20 sec
***Vtach
-30% within 20 sec
>110 bpm for 10 sec
LV Shock left ventricular shock ARTsys
CVPmean
PAPdia
Perf
<78 mmHg for 300 sec
<16 mmHg for 300 sec
>16 mmHg for 300 sec
<1.2 for 300 sec
TPX & TPND tamponade (obstructive shock) ARTsys
CVPmean
Perf
PAPdia
<78 mmHg for 180 sec
>16 mmHg for 180 sec
-20% within 3 min
>16 mmHg for 180 sec
Hypovl hypovolemia ARTmean
CVP
Perf
NIBPm
<50 mmHg for 300 sec
<5 mmHg for 300 sec
-20% within 120 sec/10 min
<55 mmHg for 300 sec

1. SVT + BP – Supraventricular Tachycardia and Blood Pressure – This is intended to indicate high heart rate with low blood pressure, as frequently occurs in patients with Atrial fibrillation and a rapid ventricular rate. Tachycardia associated with hypertension, as commonly occurs with light sedation, would not trigger this alarm.

2. VTACH + BP – This is intended to indicate ventricular tachycardia with low blood pressure. This definition would be much less likely to be triggered by motion artifact than the EKG alarm is.

3. LV SHOCK – This is intended to detect Left ventricular failure (cardiogenic shock).

4. TPX & TPND – This is intended to detect either tamponade or tension pneumothorax.

5. HYPOVL – This is intended to indicate low blood pressure from hypovolemia.