. 2013 Sep;81(9):3395–3405. doi: 10.1128/IAI.00420-13

Table 3.

Experimental parameters for LA-ICP-MS

Assay and parametera Value
Laser ablation system
    Wavelength of Nd:YAG laser (nm) 200
    Laser spot size (μm) 100
    Distance between lines (μm) 10
    Laser energy (%) 50
    Laser shot frequency (Hz) 20
    Scan rate (μm/s) 20
ICP-MS
    RF power (W) 1,250
    Cool gas (Ar) (liters min−1) 16.00
    Auxiliary gas (Ar) (liters min−1) 0.8
    Sample gas (Ar) (liters min−1) 1.0
    Resolution mode Medium (4,300)
    Isotopes measured 44Ca, 55Mn, 66Zn
    Samples per peak (no.) 20
a

The laser ablation system used was LSX-213, CETAC. For ICP-MS, an Element 2 device (Thermo Fisher Scientific) was used.