Skip to main content
. 2013 Aug;20(8):1203–1212. doi: 10.1128/CVI.00089-13

Table 3.

Comparison of IgM test results obtained by the Vidas system with results obtained by the Architect and the Liaison systems in the four groups of sera included in the study

System Result No. of samples with indicated Vidas result
Positive Gray zone Negative Total
Architect Positive 92 1 0 93
Gray zone 3 1 0 4
Negative 6 2 582 590
Total 101 4 582 687
Liaison Positive 84 1 4 89
Gray zone 8 1 9 18
Negative 8 2 566 576
Total 100 4 579 683