Table 1.

Sensitivity, specificity, percent agreement, and kappa value for InBios DENV detect IgM capture ELISA compared to the in-house MAC-ELISA

Result by InBios DENV Detect IgM capture ELISA (n = 79)b No. of samples with result by in-house MAC-ELISAc
% (95% CI)
Kappa value
Positive Negative Presumptive positive Sensitivity Specificity Agreement
Positive 22 2 0
Negative 0 50 0
Equivocala 2 1 2
Total 24 53 2 92 (73–99) 94 (84–99) 94 (86–98) 0.87
a

Equivocal samples were considered false negative for sensitivity and false positive for specificity.

b

InBios IgM positive, ISR of >2.84; InBios IgM equivocal, ISR of 1.65 to 2.84; InBios IgM negative, ISR of <1.65.

c

MAC-ELISA positive, P/N of ≥3; MAC-ELISA presumptive positive, P/N of ≥2 and <3; MAC-ELISA negative, P/N of <2.