Skip to main content
. 2013 Jul 19;13(7):9267–9293. doi: 10.3390/s130709267

Table 2.

Optimal parameters that were determined for detecting green fluorescent protein (GFP) emission in mixed samples with high autofluorescence.

Experiment/Parameter Widefield Microscope Confocal Microscope
Confocal pinhole NA 4.9 Airy disk units
Illumination power Arc lamp–100% 405 laser–60% 57 laser–2%
Detector gain EMCCD–3900 PMT–138
Photobleaching Similar bleaching rates in widefield and confocal for AF. Hoechst and GFP-Less Hoechst and GFP comparatively high
SNR Low SNR due to high background signal (SNR:2–3) Medium SNR (SNR: 3–10)
RMS error High Low
Sensitivity Linear: 0.929 slope (GFP detected/GFP added) Linear: 0.96 slope (GFP detected/ GFP added)
Specificity 0.06% Type I errors 0.16 % Type I errors