Skip to main content
. Author manuscript; available in PMC: 2014 Sep 1.
Published in final edited form as: J Med Virol. 2013 Jun 17;85(9):1661–1668. doi: 10.1002/jmv.23641

TABLE III.

Comparison experiments to determine the stool detection limit of ELISA for rotavirus vs. real time PCR quantitation of the VP6 gene

Standard plasmid copy Cq values FFU/ml (WC3 vaccine virus strain) Cq values Stool detection ELISA results
7.90E+07 11.85 1.00E+07 12.04 Positive
7.90E+06 15.76 1.00E+06 16.96 Positive
7.90E+05 19.24 1.00E+05 19.77 Positive
7.90E+04 24.2 1.00E+04 23.52 Negative
7.90E+03 28.97 1.00E+03 29.8 Negative
7.90E+02 32.84 1.00E+02 33.01 Negative