Skip to main content
. Author manuscript; available in PMC: 2014 Apr 1.
Published in final edited form as: Ultrasound Med Biol. 2013 Feb 13;39(4):721–734. doi: 10.1016/j.ultrasmedbio.2012.11.019

Table 3.

Results of KLM modeling, 1.2 MHz excitation response

Without skull With skull (within window) With skull (outside of window)
Pulse-echo amplitude relative to no skull case 0 dB(reference) −23.08 dB −33.32 dB
Center frequency (f0) 1.08 MHz 1.07 MHz 1.00 MHz
−6 dB bandwidth 32.89% 28.31% 16.61%