Skip to main content
. 2013 Sep 3;9:172. doi: 10.1186/1746-6148-9-172

Table 4.

Intra-assay performance of the TaqMan probe based real time PCR assays

 
Reaction 1
Reaction 2
Reaction 3
Theoretical copy number Average Ct Average calculated copy number CV % Average Ct Average calculated copy number CV % Average Ct Average calculated copy number CV %
MVV assay
10 6
20
853389
8.67%
20.04
1008108
23.29%
19.83
1007657
7.96%
10 5
23.29
105939
9.25%
23.40
102693
9.62%
23.32
97177
9.14%
10 4
26.69
12190
3.14%
26.66
11365
1.43%
26.64
10483
7.21%
10 3
30.29
1248
6.60%
30.07
1136
5.75%
30.16
990
7.14%
10 2
34.65
83
37.84%
34.01
79
1.35%
33.59
103
34.05%
CAEV assay
10 5
21.22
87689
38.76
20.96
94514
31.38
21.71
92044
30.09
10 4
23.55
17265
18.81
24.41
11326
17.53
24.91
11230
16.01
10 3
27.33
1463
13.65
28.16
1160
20.64
27.94
1193
3.18
10 2
31.15
66
39.76
32.58
79
17.85
33.06
55
1.84
10 34.4 11 46.69 35.73 13 58.26 35.56 11 2.3