Skip to main content
. 2013 Sep 2;84(4):437–441. doi: 10.3109/17453674.2013.823590

Figure 5.

Figure 5.

Material present on the trunnion of the femoral component (panel A) was investigated by EDAX, revealing chromium (Cr), molybdenum (Mo), and oxygen (O) peaks consistent with corrosion products (panel B). The organic material within the grooves containing chromium oxides is shown at approx. 30× (panel C) and 500× magnification (panel D).