Skip to main content
. Author manuscript; available in PMC: 2013 Sep 10.
Published in final edited form as: Biomed Chromatogr. 2010 Sep 15;25(5):628–634. doi: 10.1002/bmc.1498

Table 1.

Calibration data resulting from linear least squares regression analysis for the determination of BA-TPQ in plasma and various tissue samples

Linear
equation
Slope
(±SD)
Intercept
(±SD)
Determination
coefficient (R)
Linear range
(ng/mL)
LOD
(ng/mL)
LOQ
(ng/mL)
Plasma Y=1379035.62
×
+ 10948.50
1379035.62 ±
14525.26
10948.50 ±
6450.02
0.9999 3.910–1955 2.19 3.91
Liver Y= 414597.29
× −3054.70
414597.29 ±
14602.89
−3054.70 ±
6528.10
0.9977 19.55–1955 7.82 19.55
Brain Y= 579717.02
× − 12593.05
579717.02±
11423.93
−12593.05 ±
1074.38
0.9945 19.55–1955 7.82 19.55
Spleen Y= 440809.45
× + 4942.39
440809.45 ±
29347.48
4942.39 ±
1823.16
0.9948 19.55–1955 7.82 19.55
Heart Y= 489626.21
× − 9316.55
489626.21±
11727.36
−9316.55 ±
3075.53
0.9961 19.55–3910 7.82 19.55
Lung Y=516549.20
× − 13337.27
516549.20 ±
4750.98
−13337.27 ±
3433.46
0.9960 19.55–3910 7.82 19.55
Kidney Y= 440809.45
×− 13151.46
440809.45±
29347.4768
−13151.46 ±
520.35
0.9984 19.55–3910 7.82 19.55