2 × 2 contingency tables used to calculate test operating characteristics for tceMEP, peroneal nerve SSEP, and spontaneous EMG monitoring

tceMEP No deficit Deficit
Positive 6 5
Negative 38 0
Sensitivity = 100%
Specificity = 86%
SSEP No deficit Deficit
Positive 2 3
Negative 30 1
Sensitivity = 75%
Specificity = 94%
EMG No deficit Deficit
Positive 3 1
Negative 40 4
Sensitivity = 20%
Specificity = 93%