Table 1. Values of the constant A for the line F=Ax.
Speculum | Stiffness (N/mm) | R2 |
---|---|---|
Disposable 1 | 0.0871 | 0.992 |
Disposable 2 | 0.1037 | 0.992 |
Disposable 3 | 0.0968 | 0.996 |
Disposable 4 | 0.0988 | 0.996 |
Reusable 1 | 0.1375 | 0.987 |
Reusable 2 | 0.1327 | 0.997 |
Reusable 3 | 0.1325 | 0.997 |
Reusable 4 | 0.1364 | 0.997 |
R2 is a squared correlation coefficient and shows how well the line fits the data points.