Table 5.
Out-of-field measurements for a 15×15 cm2 18 MV beam in a phantom as a function of distance from the radiation field edge. Bolus thickness: 3.5 cm, MU: 300 MUs, SSD: 100 cm.
| Distance from field edge (cm) |
Ionization chamber (cGy) |
KCl:Eu2+ (in-field calibration) (cGy) |
KCl:Eu2+ (outside-field calibration) (cGy) |
Error (cGy) |
|---|---|---|---|---|
| 10 | 4.80 | 6.21±0.27 | 4.80±0.21 | N/A |
| 20 | 1.61 | 2.22±0.16 | 1.72±0.12 | 0.11 |
| 30 | 0.67 | 0.97±0.06 | 0.75±0.05 | 0.08 |
| 40 | 0.42 | 0.69±0.03 | 0.53±0.02 | 0.11 |
| 48.5 | 0.25 | 0.39±0.01 | 0.30±0.01 | 0.05 |