Published in final edited form as: Clin Neurophysiol. 2011 Sep 21;123(4):670–680. doi: 10.1016/j.clinph.2011.07.050

Table 1.

Subject level detector performance. Overall results for S86 and S95 operating points and patient to patient results at S95. The given values correspond to the threshold levels TH A and TH B ([μV]) of the classification stage, number of expert’s marks (Nexpert) and automatic detections (Nauto), sensitivity (S), false positive rate per minute (FPR), positive predicted (PP) and positive agreement (PA).

Patient TH A TH B Nexpert Nauto S (%) FPR PP (%) PA (%)
S86 1284 1755 85.6 1.6 57.3 69.2
S95 1284 2794 94.9 3.6 39.7 57.0
A 2.97 1.34 12 69 83.3 2.0 14.5 24.7
B 3.03 1.35 12 162 100 5.0 7.4 13.8
C 3.01 1.34 37 82 97.3 1.5 43.9 60.5
D 3.03 1.34 41 175 100 4.2 23.4 38.0
E 3.38 1.40 153 212 93.5 2.4 65.6 77.3
F 3.38 1.44 74 136 85.1 2.4 46.3 60.0
G 3.03 1.34 55 203 98 4.2 27.1 42.2
H 3.38 1.34 169 390 99 7.4 43.3 60.3
I 2.85 1.34 241 303 92.5 4.0 60.7 74.8
J 2.97 1.34 3 170 33.3 5.0 0.6 1.2
K 3.01 1.35 42 101 90.5 2.1 36.6 52.4
L 3.03 1.34 26 129 100 3.5 20.2 33.5
M 3.13 1.34 212 288 96.7 4.5 47.9 68.6
N 3.13 1.34 84 226 98.8 4.3 35.4 52.6
O 2.97 1.34 123 148 93.5 1.0 79.1 85.6