Table 4.

Precision and accuracy of quality control samples

Analyte Nominal concentration (ng/mL) Accuracy (%bias) Precision (%CV) Inter-day
Day 1 Day 2 Day 3 Day 1 Day 2 Day 3
8-iso-PGF2α 5 0.68 −7.55 0.74 1.94 3.94 5.76 1.59
0.5 9.72 −5.40 −1.11 4.74 4.78 6.29 0.90
0.07 14.46 −0.77 3.43 6.16 6.49 9.13 4.15
8-iso-15(R)-PGF2α 5 −2.56 −3.33 −1.90 2.98 0.92 4.67 3.22
0.5 2.40 −2.23 3.27 6.45 8.49 7.53 1.15
0.07 −3.68 −6.04 3.07 16.59 14.59 3.88 8.29
PGF2α 5 −1.28 −6.48 −1.67 1.56 6.49 2.84 2.40
0.5 8.28 6.06 −1.67 5.98 6.75 8.95 1.10
0.07 0.10 −12.02 16.24 12.18 12.71 15.41 8.88
15(R)-PGF2α 5 −0.32 −2.89 −9.47 2.45 5.79 4.84 1.67
0.5 4.16 12.25 0.78 4.64 6.11 3.93 1.90
0.07 −11.54 19.46 −8.23 8.97 9.68 3.12 9.12

Inter-day variation (15 replicates at each concentration)