Table 2. Results in axial pullout tests in three groups (n = 8, means ± SD).
Study periods | Parameters | CPS group | PMMA-PS group | EPS group |
6-week | Fmax (N) | 827.88±139.22 | 1426.38±235.75* | 1252.13±203.51*# |
E (J) | 1.66±0.30 | 2.84±0.55* | 2.48±0.45*# | |
12-week | Fmax (N) | 906.63±152.50 & | 1472.75±248.65*& | 1518.88±256.81*#Δ |
E (J) | 1.80±0.35 & | 2.95±0.60*& | 3.09±0.59*#Δ |
CPS, PMMA-PS and EPS indicate conventional pedicle screw, polymethylmethacrylate-augmented pedicle screw and expansive pedicle screw, respectively. Fmax and E indicate the maximum pullout strength and energy-to-failure, respectively.
Compared with CPS group at the same study period, p<0.05; # Compared with PMMA-PS group at the same study period, p>0.05; & Compared with 6-week in the same group, p>0.05; Δ Compared with 6-week in the same group, p<0.05.