. 2013 Apr 16;26(5):971–976. doi: 10.1007/s10278-013-9602-y

Table 5.

Comparison of image quality scores and processing times between the CADstream and the DynaCAD for breast systems

Score categories Number of observations Mean Std Min Max p value
Maximum intensity projection score
CS 163 3.05 0.58 1 4 0.58
DC 172 2.97 0.94 0 4
Reformat score
CS 163 2.38 1.08 0 4 0.05
DC 171 2.26 1.16 0 4
Motion correction score
CS 163 2.26 1.26 0 4 <0.0001
DC 172 1.29 1.19 0 3
Color parametric map score
CS 163 3.14 0.55 1 4 0.24
DC 172 3.08 0.5 0 4
Time–intensity curve score
CS 163 3.09 0.51 0 4 0.59
DC 172 3.06 0.47 0 4
Volume rendering score
CS 163 2.94 1.32 0 5 <0.0001
DC 172 1.28 1.54 0 5
Overall score
CS 165 3.16 0.74 1 5 0.08
DC 172 3.04 0.31 2 4
Reader interpretation time
CS 163 12.69 4.55 5 45 0.11
DC 172 11.98 2.98 6 20

Std standard deviation, Min minimum, Max maximum, CS CADstream, DC DynaCAD for Breast

The maximum number of observations is 177. p values are based on signed-rank test. Using the Bonferroni correction for multiple comparisons (six comparisons for image quality), the threshold for significance at a 5 % level is 0.05/9 = 0.006